FEATURES
* Reads through motion and low perfusion
conditions
* Screen rotation provides upright display
or vertical or horizontal monitor positioning
* SpO2, pulse rate, alarm, plethymographic
waveforms display and up to 72 hours
of patient trends.
* User configurability maximizes operation
for all applications
* RS-232 output (printer or communication)
* Designed for desktop and handheld
* Monitor adult. pediatrics and neonates
* Protect ESU (electro surgical unit) noise
* Nellcor compatable probe
PERFORMANCS
Measurement range
Pulse rate : 25 to 250bpm
%SpO2 :0to100%
Perfusion :0.1 to 20%
Saturation accuracy
Without motion Adult: 70% to 100% ±2 digits
Neonates : 70% to 100% ±3digits
With motion Adult : 70% to 100% ±3 digits
Neonates : 70% to 100% ±4 digits
Pulse rate accuracy
Without motion 25 to 250bpm ±3 digits
With motion 25 to 250bpm ±5 digits
ELECTRICAL
Instrument
AC Power requirements : 100 to 240VAC, 50/60Hz
Power Consumption : 24VA
Fuse Rating : 1 Amp, 250V
Battery
Type : Li-ion
Capacity : up to 2hours
Charging time : 5hours
ENVIRONMENTAL
- Operating Temperature : 5°C to 40°c
- Storage Temperature : -20°C to 70°C
- Operating Humidity : 5% to 95%
- Operating Altitude : -1,000 ft to 12,000 ft
PHYSICAL CHARACTERISTICS
Weight : main 650g, adaptor 550g
Dimensions : 9.6" x 1.6" x 3.3" (24.5cm x 4.2cm x 8.5cm)
Equipment Classfication - IEC 60601-1 / EN865 / EN/IEC 60601-1-2
Alarms - Audible and visual alarms for high and low saturation and pulse rate
- sensor condition, system failure and low battery
Display/Indicators
- Type : yellow / green LCD - Pixels : 240 x 64 dots - Dot Pitch : 0.53mm
- Data display : %SpO2, pulse rate, %SpO2 limits, pulse rate limits
pleth waveform, alarm status, trends, Battery status, time
Output Interface - Serial RS-232
Trend Memory
- Saturation and Pulse rate Memories ( 72hours maximum )
Standard Accessories
- IV pole mounting brackets / Reusable adult spo2 sensor
Option
- External thermal printer / Blue LCD |